Polysomnography is a gold standard for a sleep study, and it provides very accurate results, but its cost (both personnel and material) are quite high. Therefore, the development of a low-cost system for overnight breathing and heartbeat monitoring, which provides more comfort while recording the data, is a well-motivated challenge. The system proposed in this manuscript is based on the usage of resistive pressure sensors installed under the mattress. These sensors can measure slight pressure changes provoked during breathing and heartbeat. The captured signal requires advanced processing, like applying filters and amplifiers before the analog signal is ready for the next step. Then, the output signal is digitalized and further processed by an algorithm that performs a custom filtering before it can recognize breathing and heart rate in real-time. The result can be directly visualized. Furthermore, a CSV file is created containing the raw data, timestamps, and unique IDs to facilitate further processing. The achieved results are promising, and the average deviation from a reference device is about 4bpm.
Objective: This paper presents an algorithm for non-invasive sleep stage identification using respiratory, heart rate and movement signals. The algorithm is part of a system suitable for long-term monitoring in a home environment, which should support experts analysing sleep. Approach: As there is a strong correlation between bio-vital signals and sleep stages, multinomial logistic regression was chosen for categorical distribution of sleep stages. Several derived parameters of three signals (respiratory, heart rate and movement) are input for the proposed method. Sleep recordings of five subjects were used for the training of a machine learning model and 30 overnight recordings collected from 30 individuals with about 27 000 epochs of 30 s intervals each were evaluated. Main results: The achieved rate of accuracy is 72% for Wake, NREM, REM (with Cohen's kappa value 0.67) and 58% for Wake, Light (N1 and N2), Deep (N3) and REM stages (Cohen's kappa is 0.50). Our approach has confirmed the potential of this method and disclosed several ways for its improvement. Significance: The results indicate that respiratory, heart rate and movement signals can be used for sleep studies with a reasonable level of accuracy. These inputs can be obtained in a non-invasive way applying it in a home environment. The proposed system introduces a convenient approach for a long-term monitoring system which could support sleep laboratories. The algorithm which was developed allows for an easy adjustment of input parameters that depend on available signals and for this reason could also be used with various hardware systems.