Accurate monitoring of a patient's heart rate is a key element in the medical observation and health monitoring. In particular, its importance extends to the identification of sleep-related disorders. Various methods have been established that involve sensor-based recording of physiological signals followed by automated examination and analysis. This study attempts to evaluate the efficacy of a non-invasive HR monitoring framework based on an accelerometer sensor specifically during sleep. To achieve this goal, the motion induced by thoracic movements during cardiac contractions is captured by a device installed under the mattress. Signal filtering techniques and heart rate estimation using the symlets6 wavelet are part of the implemented computational framework described in this article. Subsequent analysis indicates the potential applicability of this system in the prognostic domain, with an average error margin of approximately 3 beats per minute. The results obtained represent a promising advancement in non-invasive heart rate monitoring during sleep, with potential implications for improved diagnosis and management of cardiovascular and sleep-related disorders.
Sleep quality and in general, behavior in bed can be detected using a sleep state analysis. These results can help a subject to regulate sleep and recognize different sleeping disorders. In this work, a sensor grid for pressure and movement detection supporting sleep phase analysis is proposed. In comparison to the leading standard measuring system, which is Polysomnography (PSG), the system proposed in this project is a non-invasive sleep monitoring device. For continuous analysis or home use, the PSG or wearable Actigraphy devices tends to be uncomfortable. Besides this fact, they are also very expensive. The system represented in this work classifies respiration and body movement with only one type of sensor and also in a non-invasive way. The sensor used is a pressure sensor. This sensor is low cost and can be used for commercial proposes. The system was tested by carrying out an experiment that recorded the sleep process of a subject. These recordings showed the potential for classification of breathing rate and body movements. Although previous researches show the use of pressure sensors in recognizing posture and breathing, they have been mostly used by positioning the sensors between the mattress and bedsheet. This project however, shows an innovative way to position the sensors under the mattress.
This document presents a new complete standalone system for a recognition of sleep apnea using signals from the pressure sensors placed under the mattress. The developed hardware part of the system is tuned to filter and to amplify the signal. Its software part performs more accurate signal filtering and identification of apnea events. The overall achieved accuracy of the recognition of apnea occurrence is 91%, with the average measured recognition delay of about 15 seconds, which confirms the suitability of the proposed method for future employment. The main aim of the presented approach is the support of the healthcare system with the cost-efficient tool for recognition of sleep apnea in the home environment.
In many cases continuous monitoring of vital signals is required and low intrusiveness is an important requirement. Incorporating monitoring systems in the hospital or home bed could have benefits for patients and caregivers. The objective of this work is the definition of a measurement protocol and the creation of a data set of measurements using commercial and low-cost prototypes devices to estimate heart rate and breathing rate. The experimental data will be used to compare results achieved by the devices and to develop algorithms for feature extraction of vital signals.
The scoring of sleep stages is an essential part of sleep studies. The main objective of this research is to provide an algorithm for the automatic classification of sleep stages using signals that may be obtained in a non-obtrusive way. After reviewing the relevant research, the authors selected a multinomial logistic regression as the basis for their approach. Several parameters were derived from movement and breathing signals, and their combinations were investigated to develop an accurate and stable algorithm. The algorithm was implemented to produce successful results: the accuracy of the recognition of Wake/NREM/REM stages is equal to 73%, with Cohen's kappa of 0.44 for the analyzed 19324 sleep epochs of 30 seconds each. This approach has the advantage of using the only movement and breathing signals, which can be recorded with less effort than heart or brainwave signals, and requiring only four derived parameters for the calculations. Therefore, the new system is a significant improvement for non-obtrusive sleep stage identification compared to existing approaches.
The importance of sleep for human life is enormous. It affects physical, mental, and psychological health. Therefore, it is vital to recognise sleep disorders in a timely manner in order to be able to initiate therapy. There are two methods for measuring sleep-related parameters - objective and subjective. Whether the substitution of a subjective method for an objective one is possible is investigated in this paper. Such replacement may bring several advantages, including increased comfort for the user. To answer this research question, a study was conducted in which 75 overnight recordings were evaluated. The primary purpose of this study was to compare both ways of measurement for total sleep time and sleep efficiency, which are essential parameters for, e.g., insomnia diagnosis and treatment. The evaluation results demonstrated that, on average, there are 32 minutes of difference between the two measurement methods when total sleep time is analysed. In contrast, on average, both measurement methods differ by 7.5% for sleep efficiency measurement. It should also be noted that people typically overestimate total sleep time and efficiency with the subjective method, where the perceived values are measured.
Health monitoring in a home environment can have broader use since it may provide continuous control of health parameters with relatively minor intrusiveness into regular life. This work aims to verify if it is possible to replace the typical in some sleep medicine areas subjective questioning by an objective measurement using electronic devices. For this purpose, a study was conducted with ten subjects, in which objective and subjective measurement of relevant sleep parameters took place. The results of both measurement methods were evaluated and analyzed. The results showed that while for some measures, such as Total Time in Bed, there is a high agreement between objective and subjective measurements, for others, such as sleep quality, there are significant differences. For this reason, currently, a combination of both measurement methods may be beneficial and provide the most detailed results, while a partial replacement can already reduce the number of questions at the subjective measurement by measurement through electronic devices.
The influence of sleep on human health is enormous. Accordingly, sleep disorders can have a negative impact on it. To avoid this, they should be identified and treated in time. For this purpose, objective (with an appropriate device) or subjective (based on perceived values) measurement methods are used for sleep analysis to understand the problem. The aim of this work is to find out whether an exchange of the two methods is possible and can provide reliable results. In accordance with this goal, a study was conducted with people aged over 65 years old (a total of 154 night-time recordings) in which both measurement methods were compared. Sleep questionnaires and electronic devices for sleep assessment placed under the mattress were applied to achieve the study aims. The obtained results indicated that the correlation between both measurement methods could be observed for sleep characteristics such as total sleep time, total time in bed and sleep efficiency. However, there are also significant differences in absolute values of the two measurement approaches for some subjects/nights, which leads us to conclude that the substitution is more likely to be considered in case of long-term monitoring where the trends are of more importance and not the absolute values for individual nights.
The recovery of our body and brain from fatigue directly depends on the quality of sleep, which can be determined from the results of a sleep study. The classification of sleep stages is the first step of this study and includes the measurement of vital data and their further processing. The non-invasive sleep analysis system is based on a hardware sensor network of 24 pressure sensors providing sleep phase detection. The pressure sensors are connected to an energy-efficient microcontroller via a system-wide bus. A significant difference between this system and other approaches is the innovative way in which the sensors are placed under the mattress. This feature facilitates the continuous use of the system without any noticeable influence on the sleeping person. The system was tested by conducting experiments that recorded the sleep of various healthy young people. Results indicate the potential to capture respiratory rate and body movement.