The respiratory rate is a vital sign indicating breathing illness. It is necessary to analyze the mechanical oscillations of the patient's body arising from chest movements. An inappropriate holder on which the sensor is mounted, or an inappropriate sensor position is some of the external factors which should be minimized during signal registration. This paper considers using a non-invasive device placed under the bed mattress and evaluates the respiratory rate. The aim of the work is the development of an accelerometer sensor holder for this system. The normal and deep breathing signals were analyzed, corresponding to the relaxed state and when taking deep breaths. The evaluation criterion for the holder's model is its influence on the patient's respiratory signal amplitude for each state. As a result, we offer a non-invasive system of respiratory rate detection, including the mechanical component providing the most accurate values of mentioned respiratory rate.
This paper presents a bed system able to analyze a person’s movement, breathing and recognize the positions that the subject is lying on the bed during the night without any additional physical contact. The measurements are performed with sensors placed between the mattress and the bed-frame. An Intel Edison board was used as an endpoint that served as a communication node from the mesh network to external service. Two nodes and Intel Edison are attached to the bottom of the bed frame and they are connected to the sensors. First test results have indicated the potential of the proposed approach for the recognition of sleep positions with 83% of correct recognized positions.
The process of restoring our body and brain from fatigue is directly depend-ing on the quality of sleep. It can be determined from the report of the sleep study results. Classification of sleep stages is the first step of this study and this includes the measurement of biovital data and its further processing.
In this work, the sleep analysis system is based on a hardware sensor net, namely a grid of 24 pressure sensors, supporting sleep phase recognition. In comparison to the leading standard, which is polysomnography, the proposed approach is a non-invasive system. It recognises respiration and body move-ment with only one type of low-cost pressure sensors forming a mesh archi-tecture. The nodes implement as a series of pressure sensors connected to a low-power and performant microcontroller. All nodes are connected via a system wide bus with address arbitration. The embedded processor is the mesh network endpoint that enables network configuration, storing and pre-processing of the data, external data access and visualization.
The system was tested by executing experiments recording the sleep of different healthy young subjects. The results obtained have indicated the po-tential to detect breathing rate and body movement. A major difference of this system in comparison to other approaches is the innovative way to place the sensors under the mattress. This characteristic facilitates the continuous using of the system without any influence on the common sleep process.
Methods based exclusively on heart rate hardly allow to differentiate between physical activity, stress, relaxation, and rest, that is why an additional sensor like activity/movement sensor added for detection and classification. The response of the heart to physical activity, stress, relaxation, and no activity can be very similar. In this study, we can observe the influence of induced stress and analyze which metrics could be considered for its detection. The changes in the Root Mean Square of the Successive Differences provide us with information about physiological changes. A set of measurements collecting the RR intervals was taken. The intervals are used as a parameter to distinguish four different stages. Parameters like skin conductivity or skin temperature were not used because the main aim is to maintain a minimum number of sensors and devices and thereby to increase the wearability in the future.
We consider the problem of increasing the informative value of electrocardiographic (ECG) surveys using data from multichannel electrocardiographic leads, that include both recorded electrocardiosignals and the coordinates of the electrodes placed on the surface of the human torso. In this area, we were interested in reconstruction of the surface distribution of the equivalent sources during the cardiac cycle at relatively low hardware cost. In our work, we propose to reconstruct the equivalent electrical sources by numerical methods, based on integral connection between the density of electrical sources and potential in a conductive medium. We consider maps of distributions of equivalent electric sources on the heart surface (HSSM), presenting source distributions in the form of a simple or double electrical layer. We indicate the dynamics of the heart electrical activity by the space-time mapping of equivalent electrical sources in HSSM.
The recovery of our body and brain from fatigue directly depends on the quality of sleep, which can be determined from the results of a sleep study. The classification of sleep stages is the first step of this study and includes the measurement of vital data and their further processing. The non-invasive sleep analysis system is based on a hardware sensor network of 24 pressure sensors providing sleep phase detection. The pressure sensors are connected to an energy-efficient microcontroller via a system-wide bus. A significant difference between this system and other approaches is the innovative way in which the sensors are placed under the mattress. This feature facilitates the continuous use of the system without any noticeable influence on the sleeping person. The system was tested by conducting experiments that recorded the sleep of various healthy young people. Results indicate the potential to capture respiratory rate and body movement.
This document presents an algorithm for a non-obtrusive recognition of Sleep/Wake states using signals derived from ECG, respiration, and body movement captured while lying in a bed. As a core mathematical base of system data analytics, multinomial logistic regression techniques were chosen. Derived parameters of the three signals are used as the input for the proposed method. The overall achieved accuracy rate is 84% for Wake/Sleep stages, with Cohen’s kappa value 0.46. The presented algorithm should support experts in analyzing sleep quality in more detail. The results confirm the potential of this method and disclose several ways for its improvement.
With the advancement in sensor technology and the trend shift of health measurement from treatment after diagnosis to abnormalities detection long before the occurrence, the approach of turning private spaces into diagnostic spaces has gained much attention. In this work, we designed and implemented a low-cost and compact form factor module that can be deployed on the steering wheel of cars as well as most frequently touch objects at home in order to measure physiological signals from the fingertip of the subject as well as environmental parameters. We estimated the heart rate and SpO2 with the error of 2.83 bpm and 3.52%, respectively. The signal evaluation of skin temperature shows a promising output with respect to environmental recalibration. In addition, the electrodermal activity sensor followed the reference signal, appropriately which indicates the potential for further development and application in stress measurement.
Preliminary results of homomorphic deconvolution application to surface EMG signals during walking
(2021)
Homomorphic deconvolution is applied to sEMG signals recorded during walking. Gastrocnemius lateralis and tibialis anterior signals were acquired according to SENIAM recommendation. MUAP parameters like amplitude and scale were estimated, whilst the MUAP shape parameter was fixed. This features a useful time-frequency representation of sEMG signal. Estimation of scale MUAP parameter was verified extracting the mean frequency of filtered EMG signal, extracted from the scale parameter estimated with two different MUAP shape values.