Sleep disorders can impact daily life, affecting physical, emotional, and cognitive well-being. Due to the time-consuming, highly obtrusive, and expensive nature of using the standard approaches such as polysomnography, it is of great interest to develop a noninvasive and unobtrusive in-home sleep monitoring system that can reliably and accurately measure cardiorespiratory parameters while causing minimal discomfort to the user’s sleep. We developed a low-cost Out of Center Sleep Testing (OCST) system with low complexity to measure cardiorespiratory parameters. We tested and validated two force-sensitive resistor strip sensors under the bed mattress covering the thoracic and abdominal regions. Twenty subjects were recruited, including 12 males and 8 females. The ballistocardiogram signal was processed using the 4th smooth level of the discrete wavelet transform and the 2nd order of the Butterworth bandpass filter to measure the heart rate and respiration rate, respectively. We reached a total error (concerning the reference sensors) of 3.24 beats per minute and 2.32 rates for heart rate and respiration rate, respectively. For males and females, heart rate errors were 3.47 and 2.68, and respiration rate errors were 2.32 and 2.33, respectively. We developed and verified the reliability and applicability of the system. It showed a minor dependency on sleeping positions, one of the major cumbersome sleep measurements. We identified the sensor under the thoracic region as the optimal configuration for cardiorespiratory measurement. Although testing the system with healthy subjects and regular patterns of cardiorespiratory parameters showed promising results, further investigation is required with the bandwidth frequency and validation of the system with larger groups of subjects, including patients.
The digital twin concept has been widely known for asset monitoring in the industry for a long time. A clear example is the automotive industry. Recently, there has also been significant interest in the application of digital twins in healthcare, especially in genomics in what is known as precision medicine. This work focuses on another medical speciality where digital twins can be applied, sleep medicine. However, there is still great controversy about the fundamentals that constitute digital twins, such as what this concept is based on and how it can be included in healthcare effectively and sustainably. This article reviews digital twins and their role so far in what is known as personalized medicine. In addition, a series of steps will be exposed for a possible implementation of a digital twin for a patient suffering from sleep disorders. For this, artificial intelligence techniques, clinical data management, and possible solutions for explaining the results derived from artificial intelligence models will be addressed.
Personalized remote healthcare monitoring is in continuous development due to the technology improvements of sensors and wearable electronic systems. A state of the art of research works on wearable sensors for healthcare applications is presented in this work. Furthermore, a state of the art of wearable devices, chest and wrist band and smartwatches available on the market for health and sport monitoring is presented in this paper. Many activity trackers are commercially available. The prices are continuously reducing and the performances are improving, but commercial devices do not provide raw data and are therefore not useful for research purposes.
This paper compares two popular scripting implementations for hardware prototyping: Python scripts exe- cut from User-Space and C-based Linux-Driver processes executed from Kernel-Space, which can provide information to researchers when considering one or another in their implementations. Conclusions exhibit that deploying software scripts in the kernel space makes it possible to grant a certain quality of sensor information using a Raspberry Pi without the need for advanced real-time operational systems.
The respiratory rate is a vital sign indicating breathing illness. It is necessary to analyze the mechanical oscillations of the patient's body arising from chest movements. An inappropriate holder on which the sensor is mounted, or an inappropriate sensor position is some of the external factors which should be minimized during signal registration. This paper considers using a non-invasive device placed under the bed mattress and evaluates the respiratory rate. The aim of the work is the development of an accelerometer sensor holder for this system. The normal and deep breathing signals were analyzed, corresponding to the relaxed state and when taking deep breaths. The evaluation criterion for the holder's model is its influence on the patient's respiratory signal amplitude for each state. As a result, we offer a non-invasive system of respiratory rate detection, including the mechanical component providing the most accurate values of mentioned respiratory rate.
Assistive environments are entering our homes faster than ever. However, there are still various barriers to be broken. One of the crucial points is a personalization of offered services and integration of assistive technologies in common objects and therefore in a regular daily routine. Recognition of sleep patterns for the preliminary sleep study is one of the health services that could be performed in an undisturbing way. This article proposes the hardware system for the measurement of bio-vital signals necessary for initial sleep study in a non-obtrusive way. The first results confirm the potential of measurement of breathing and movement signals with the proposed system.
Monitoring heart rate and breathing is essential in understanding the physiological processes for sleep analysis. Polysomnography (PSG) system have traditionally been used for sleep monitoring, but alternative methods can help to make sleep monitoring more portable in someone's home. This study conducted a series of experiments to investigate the use of pressure sensors placed under the bed as an alternative to PSG for monitoring heart rate and breathing during sleep. The following sets of experiments involved the addition of small rubber domes - transparent and black - that were glued to the pressure sensor. The resulting data were compared with the PSG system to determine the accuracy of the pressure sensor readings. The study found that the pressure sensor provided reliable data for extracting heart rate and respiration rate, with mean absolute errors (MAE) of 2.32 and 3.24 for respiration and heart rate, respectively. However, the addition of small rubber hemispheres did not significantly improve the accuracy of the readings, with MAEs of 2.3 bpm and 7.56 breaths per minute for respiration rate and heart rate, respectively. The findings of this study suggest that pressure sensors placed under the bed may serve as a viable alternative to traditional PSG systems for monitoring heart rate and breathing during sleep. These sensors provide a more comfortable and non-invasive method of sleep monitoring. However, the addition of small rubber domes did not significantly enhance the accuracy of the readings, indicating that it may not be a worthwhile addition to the pressure sensor system.
Background: Polysomnography (PSG) is the gold standard for detecting obstructive sleep apnea (OSA). However, this technique has many disadvantages when using it outside the hospital or for daily use. Portable monitors (PMs) aim to streamline the OSA detection process through deep learning (DL).
Materials and methods: We studied how to detect OSA events and calculate the apnea-hypopnea index (AHI) by using deep learning models that aim to be implemented on PMs. Several deep learning models are presented after being trained on polysomnography data from the National Sleep Research Resource (NSRR) repository. The best hyperparameters for the DL architecture are presented. In addition, emphasis is focused on model explainability techniques, concretely on Gradient-weighted Class Activation Mapping (Grad-CAM).
Results: The results for the best DL model are presented and analyzed. The interpretability of the DL model is also analyzed by studying the regions of the signals that are most relevant for the model to make the decision. The model that yields the best result is a one-dimensional convolutional neural network (1D-CNN) with 84.3% accuracy.
Conclusion: The use of PMs using machine learning techniques for detecting OSA events still has a long way to go. However, our method for developing explainable DL models demonstrates that PMs appear to be a promising alternative to PSG in the future for the detection of obstructive apnea events and the automatic calculation of AHI.
Healthy and good sleep is a prerequisite for a rested mind and body. Both form the basis for physical and mental health. Healthy sleep is hindered by sleep disorders, the medically diagnosed frequency of which increases sharply from the age of 40. This chapter describes the formal specification of an on-course practical implementation for a non-invasive system based on biomedical signal processing to support the diagnosis and treatment of sleep-related diseases. The system aims to continuously monitor vital data during sleep in a patient’s home environment over long periods by using non-invasive technologies. At the center of the development is the MORPHEUS Box (MoBo), which consists of five main conceptualizations: the MoBo core, the MoBo-HW, the MoBo algorithm, the MoBo API, and the MoBo app. These synergistic elements aim to support the diagnosis and treatment of sleep-related diseases. Although there are related developments in individual aspects concerning the system, no comparative approach is known that gives a similar scope of functionality, deployment flexibility, extensibility, or the possibility to use multiple user groups. With the specification provided in this chapter, the MORPHEUS project sets a good platform, data model, and transmission strategies to bring an innovative proposal to measure sleep quality and detect sleep diseases from non-invasive sensors.
Sleep analysis using a Polysomnography system is difficult and expensive. That is why we suggest a non-invasive and unobtrusive measurement. Very few people want the cables or devices attached to their bodies during sleep. The proposed approach is to implement a monitoring system, so the subject is not bothered. As a result, the idea is a non-invasive monitoring system based on detecting pressure distribution. This system should be able to measure the pressure differences that occur during a single heartbeat and during breathing through the mattress. The system consists of two blocks signal acquisition and signal processing. This whole technology should be economical to be affordable enough for every user. As a result, preprocessed data is obtained for further detailed analysis using different filters for heartbeat and respiration detection. In the initial stage of filtration, Butterworth filters are used.